Type 3 commands – balance output functions, Type 4 commands – clock/calendar functions – Scientech 12000 Series User Manual
Page 18

18
Type 3 Commands – Balance Output Functions:
Command Response
Purpose
SEND[CR]
See the RS-232 Balance Message Formats section.
Returns the contents of the display register.
[register number]_RCL[CR] REG
XXX:[register
contents][current units or
'MULT.'] [CR] [LF]
Sends the contents of the specified register 
in the current units if it is additive or with 
the designator 'MULT.' if it is multiplicative. 
An unflagged number is sent with no 
designator. An undefined register returns a ? 
with a [CR][LF]. 
RCL_TARE[CR]
REG 091:[TARE register contents] [current 
units][CR][LR] 
Return the TARE register's contents.
RCL_[(CONVERT)(CAL)] REG
XXX:[contents of specified
register][MULT.][CR][LF]
Returns indicated register contents, if in that 
mode. If not in that mode then a ? with a 
[CR][LF] is returned. 
RCL_PCS[CR]
REG XXX:[contents of pieces register][last FP 
units/pc][CR][LF] 
Returns pieces register contents expressed as 
last front panel units/pc, if in that mode. If 
not in that mode then a ? with a [CR][LF] is 
returned. 
[register 
number]_RCL_TARE[CR] 
REG XXX:[register contents][current units][CR][LF] Adds the specified register contents to the
TARE register.
RCL_XAVG[CR]
[average weight] [current units] AVX[CR][LF]. [2 
second delay]. [number of samples taken] 
###[CR][LF]. [2 second delay]. [standard deviation] 
[ current units] SDEV[CR][LF]. [2 second delay]. 
[relative standard deviation] COV[CR][LF]. 
Sends the results of a standard deviation run. 
The TBAR mode, if in use, is temporarily 
suspended while this process is going on. 
 
 
 
Type 4 Commands – Clock/Calendar Functions:
Command Response
Purpose
[six digit integer] 
_ENTER_100_STORE[CR] 
Sets the calendar function, if available, to the specified 
date. The date must be specified in the MMDDYY 
format and no delimiters. 
100_RCL[CR]
REG 100: MM.DD.YY DATE[CR][LF] Returns the date.
[six digit integer] 
_ENTER_101_STORE[CR] 
Sets the clock function, if available, to the specified 
time. The time must be specified in the HHMMSS 
format using military 24 hour time format. 
101_RCL[CR]
REG 101: [HH:MM:SS] 
TIME[CR][LF] 
Returns the current time, using military 24 hour time 
format, once for each request. 
[1 to 6 digit integer] 
_ENTER_89_STORE[CR] 
Sets the auto send register to the specified send time 
interval. For the auto send feature to be operational, 
you must set the auto send selection to ON (see RS-
232 Front Panel Configuration section for details). 
The format used is HHMMSS with no delimiters. For 
instance, to set the auto send interval to 5 minutes and 
25 seconds, you would send this command string: 525 
ENTER 89 STORE[CR]. Also you must have the auto 
Send selection set to ON before the auto Send will 
work (see RS-232 Front Panel Configuration Section). 
89_RCL[CR]
REG 089: [HH:MM:SS] 
INTV[CR][LF] 
Returns the auto SEND time interval currently set.
[(1)(0)]_ENTER_79_STORE[CR]
A 1 in register 79 disables the date/time stamp feature 
of any SEND. A 0 in register 79 enables the date/time 
stamp. 
 
 
 
 