Lo-z – Milwaukee Tool 2217-20 User Manual

Page 4

background image

6

7

Dial Position

Range

Resolution

Accuracy

Voltage AC

600.0mV/6.000V

60.00V/600.0V

0.1mV/0.001V/

0.01V/0.1V

±(1.0% + 3 dgt) (45~500Hz)

±(2.0% + 3 dgt) (500~1000Hz)

Voltage DC

600.0mV/6.000V

60.00V/600.0V

0.1mV/0.001V/

0.01V/0.1V

±(0.5% + 2 dgt)

Lo-Z

Lo-Z

Low Input
Impedance

600.0V

0.1V

±(2.0% + 3 dgt) DC,AC : 45~500Hz

Hz

Hz

Hertz

99.99/999.9Hz

9.999/50.00kHz

0.01Hz/0.1Hz/

0.001kHz/0.01kHz

±(0.1% + 2 dgt)

Resistance

600.0Ω

6.000/60.00/600.0kΩ

6.000/40.00MΩ

0.1Ω

0.001/0.01/0.1kΩ

±(1.0% + 5 dgt)

.001/.01MΩ

±(1.9% + 5 dgt)

Continuity

Cont Buzzer

0-600.0Ω

Buzzer sounds at 30Ω or less

Capacitance

100.0μF

1000μF

0.1μF

1μF

±(1.9% + 2 dgt)

Current AC

0.40mA-60.00mA/600.0mA

6.000A/10.00A

0.01mA/0.1mA

0.001A/0.01A

±(1.5% + 3 dgt) (45~500Hz)

Current DC

0.40mA-60.00mA/600.0mA

6.000A/10.00A

0.01mA/0.1mA

0.001A/0.01A

±(1.0% + 3 dgt)

Temperature

- 40.0ºC ~ 400.0ºC

-40.0ºF ~ 752.0ºF

0.1°C

0.2°F

±(1.0% + 10 dgt)
±(1.0% + 18 dgt)

Functions

† Temperature Range, Resolution and Accuracy are for the DMM. Temperature Probe may have different

specifi cations.

* These instruments are True-RMS sensing. All voltage and current readings are True-RMS values.
* Input impedance: Voltage DC: 10MΩ; Voltage AC: 10MΩ // less than 100pF; Lo-z: 4kΩ approximately
* Overload protection:
Voltage DC, Voltage AC, Lo-Z Voltage and Hertz: AC/DC 720V for 10 second
Current DC & Current AC: AC/DC 20A for 10 sec.
Resistance, Continuity, Capacitance and Temperature: AC/DC 600V for 10 second
* Maximum measurement time: 1 minute at 10A, rest time 20 minutes minimum
* Minimum frequency measurement is 2Hz
* Signal voltage in frequency mode: 8V-600V(RMS)
* Minimum AC Current measurement is 0.40mA
* Minimum AC Voltage measurement is 4.0mV
* For AC Voltage, AC Current and Lo-z:
Additional Accuracy by Crest Factor (C.F.):
Add 6.0% for C.F. 1.0 ~ 2.0.
Add 7.0% for C.F. 2.0 ~ 2.5.
Add 9.0% for C.F. 2.5 ~ 3.0
Max. Crest Factor: 1.6 for 6600 ~ 5000 digits
2.0 for 5000 ~ 3000 digits
3.0 for 3000 ~ 0 digits
* Measurement accuracy of square wave and truncated waveforms at 1kHz is unspecifi ed.

ASSEMBLY

WARNING

To avoid an electrical hazard,

turn the Rotary Dial to OFF and disconnect the

test leads before replacing batteries.

Loading/Changing the Batteries
Replace batteries when the Low Battery indicator

is displayed.

1. Turn Rotary Dial to OFF and disconnect the test

leads.

2. Unscrew and remove battery door.
3. Insert two (2) AA

batteries, according to
the polarity marked in
the battery compartment

4. Close the battery door

and tighten four (4)
screws securely.

OPERATION

Before Use
Confi rm the Rotary Dial is set to the correct position,
the instrument is set to the correct measurement
mode, and the Data hold function is disabled. Oth-
erwise, desired measurement cannot be made.

LCD Backlight

The LCD backlight will turn off after about 10 min-
utes of inactivity. Press the backlight button to
turn the backlight on again.

Making a Measurement

AC Voltage

DANGER

To avoid electrical shock:

Never make measurement on a circuit in

which voltage over AC600V exists.

Do not use with the Battery Cover removed.

Keep fi ngers behind the guards and away

from test lead tips during measurements.

CAUTION

Readings may fl uctuate or

be infl uenced in noisy environment.

1. Set the Dial to

position.

2. Connect the red test lead

to the VΩ terminal and the
black test lead to the COM
terminal.

3. Connect the test leads to the

circuit under test. The read-
ing is displayed.

WARNING

Only use Milwaukee test

leads with the MILWAUKEE DMM. Inspect test

leads for continuity before each use. Do not

use if the readings are high or noisy.