14.Trouble shooting
1.During venting or drying, the thump sounds. It means that the Condensed water collector
is full. You must drain the water from the collector.
2.During a cycle, a banging sounds and large steam leaks from the chamber door and
displays
E6
code. Please verify whether the door knob is properly turned and locked.
If the steam leaks at the beginning of a cycle, you need to change the seal according to
instructions at page 14.
3.When a cycle completes, there are water drops on the instruments.
You may correct the problems by:
1). Please check the drain filter.
2). Reduce the load of the instrument in the chamber, especially the porous and dressing and
other water absorbent materials.
3). You may also wait for a while at the end of a cycle
Ed
for better drying result. The
autoclave will continue heating the chamber until you open the door.