5.3

START cycle

Close the door after loaded the instruments, press START
key. 0.5S later, electrical lock will work, with “KA”
sounds. If the electrical lock did not achieve expect
position will show door fault. It will start sterilization cycle
after locked the door successfully. Look at fig 5-4.
During sterilization cycle, press MENU/C key for 3S, it will
exit cycle.If you want to stop the cycle before drying cycle,
it shall need another 3 min to do drying cycle. During this
cycle, press MENU/C key again, it will exit drying cycle. If
press MENU /C key during drying cycle between normal
cycle, it will exit cycle directly.

5.4

Sterilization end

After sterilization, the display will show as fig 5-5 and with alarming sounds,
then you can open the door, unload the I nstruments been sterilized. At mean
time, print our the sterilization process. If the printer closed, no sterilization
process printed. According to “System setup” You could know how to close
the printer.

To ensure sterilizer's on the safe side, when water level low alarm sounds. Please be advised fill water first.
When you fill the water, please don't forget to let used water out.

0.05 Bar 83.5

C

З л ¿ ª Г Е £ ³ £ ¶ £ º £ µ £ ·

Always use the tray holder to load or unload the tray with instruments in it in order to avoid burning injuries.

Caution

Caution

Fig5-4

Fig5-5

SEA Series Steam Sterilizers

Operation manual

7