beautypg.com

Т³гж 22, Appendix – LABEC AA-17S User Manual

Page 21

background image

SEA Series Steam Sterilizers

Operation manual

17

13.Appendix.

Form 1 Work mode chart

Form 2 Error code chart

1

2

3

4

5

6

7

8

Item

9

Error

Condition

sensor error

super pressure

super temperature

constant temperature

fails

block of steam pipe

failure of the door

Steam generator over

temperature

heating overtime

pressurization

overtime

o

evaluated temperature>250 c

evaluated pressure>2.6Bar

o

chamber temperature>160 c

chamber temperature >+3C<-1C

reduced pressure<0.2Bar

after 20seconds

door opens while operation

Steam generator measures>230C

pre-heating>20 min

steam input>40 min

Solution

Examine inner temperature

sensor or pressure sensor

examine outer temperature

sensor

Return parameter autoadjus

programm

Examine the steam output

valve or vacuum valve

examine the door switch

Check steam generator

temperature sensor

examine the electronic connection

of heat ring and heat stick

examine the steam input pipe

examine sensor

Dry

Clean 20min

10min

31min

Clean test

134

134

134

121

134

2.10

2.10

2.10

1.10

2.10

S

S

S

S

S

0-7min

0-7min

0-7min

0-7min

0-7min

4min

4min

4min

4min

4min

4min

5min

5min

5min

5min

5min

4min

18min

4min

16min

3.5min

Keep pressure

10min

10min

10min

10min

10min

1min

5min

5min

5min

5min

30(25)min 44(39)min

30(25)min 42(37)min

13.5min

15min

Sterilizing temperature

Sterilizing pressure

Class

Pre-heating

Vaeuuning time

Sterilizing time

Drying time

Quick drying time

Total time

Admission time

S cycle

B&Dtest

18min

S cycle

Unwrappd

cycle

Work mode

chart

Vacuum test

This manual is related to the following products: