9 service documentation, 1 overview, setting specification, tolerances – KERN & SOHN KFU 1500V30M User Manual
Page 13

KFU-IA-e-1211
13
9 Service documentation
• This chapter is only provided for a specialist of balances !
• The weighing bridges are carried out in DMS sensor technology, at
every corner there is a DMS weighing cell.
• The analogue-digital conversion takes place in the display unit. There
are also saved all balance and country specific data.
9.1 Overview, setting specification, tolerances
Test and setting specification:
Capacity
600 kg
1500 kg
Readability
200 g
500 g
Min
4 kg
10 kg
Max
600 kg
1500 kg
1/3 corner load
200 kg
500 kg
Tolerance
200 g
500 g
Verification data and tolerances acc. to OIML:
-1
-0,75
-0,5
-0,25
0
0,25
0,5
0,75
1
0
500
1000
1500
[kg]
[g]
-0,4
-0,3
-0,2
-0,1
0
0,1
0,2
0,3
0,4
0
100
200
300
400
500
600
[kg]
[g]
600 kg
1500 kg