8.5 Instant help
In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault
Possible cause
The displayed weight is permanently
changing
• Draught/air movement
• Sites with vibration.
• Weighing plate has contact with other
objects.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
The weighing result is obviously
incorrect
• No zero display when balance unloaded
• Adjustment is no longer correct.
• Great fluctuations in temperature.
• The balance is on an uneven surface.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.