Technical data, 1 indicator, 1 4-wire system – KERN & SOHN FA 6T-3M User Manual
Page 78: 2 6-wire system

Annex page 2 of 7
Annex to Test Certificate No. DK0199-R76-11.04
Issued by DELTA
•
Indication of zero
•
Semi-automatic subtractive tare
•
Acting upon significant fault
•
Weighing unstable samples
•
Real time clock (optional)
3.
Technical data
3.1
Indicator
Type
KFN-TM / KFB-TM
Accuracy class
III or IIII
Weighing range
Single-interval, multi-interval or multi-range
Maximum number of verification scale intervals (n) 6000 for single-interval
2
×3000 for multi-interval and multi-range,
however limited to 1000 for Class IIII
Minimum input voltage per VSI
1 µV
Maximum capacity of interval or range (Max
i
): n
i
× e
i
Verification scale interval, e
i
=
Max
i
/ n
i
Initial zero-setting range:
± 10 % of Max
Maximum tare effect:
100 % of Max
Fractional factor (pi)
0.5
Excitation voltage
5 VDC
Circuit for remote sense
Active, (see below)
Minimum input impedance
87 ohm
Maximum input impedance
1600 ohm
Connecting cable to load cell(s):
See Section 3.1.1
Supply voltage:
9 - 12 VDC
230 VAC using external Vac/2Vdc adapter
Operating temperature range
Min / Max = -10 °C / +40 °C
Peripheral interface(s)
See Section 4
3.1.1
Connecting cable between the indicator and the junction box for load cell(s),
if any
3.1.1.1 4-wire system
Line
4 wires, shielded
Maximum length
The certified length of the load cell cable, which shall be con-
nected directly to the indicator.
3.1.1.2 6-wire system
Line 6
wires,
shielded
Maximum length
227 m/mm
2
Maximum resistance per wire
3.8 ohm