KFN-TM:
Access to conductor plate:
•
Remove seal
•
Open display unit
•
The application of the display unit as a weighing system able to be verified
requires that the contacts of the circuit board are short-circuited with the help
of a jumper [K1]. For non verifiable weighing systems remove the jumper.
•
To adjust, short-circuit the contacts of the circuit board, using a jumper [K2].
[K1]
[K2]