KERN & SOHN FA 6T-3M User Manual
Page 27

KFB/KFN-TM-BA_IA-e-1220
27
Optical signal:
Three colour signal lights indicate whether the load is within the two tolerance limits.
The signal lights provide the following information:
+
Goods to be weighed above tolerance limit
Red signal light
glowing
Goods to be weighed within tolerance range
Green signal light
glowing
-
Goods to be weighed below tolerance limit
Red signal light
glowing
The settings for tolerance weighing may be called up either via menu block „P0 CHK“
(see chap. 8) or faster via the key combination
Settings
Press
and
at the same time in weighing mode.
Press
until the display for entering the lower limit value
appears.
Press
, current setting will be displayed.
To enter the lower limit, e. g. 1000 Kg, press the navigation
keys (See chap. 2.1.1); the currently enabled digit will be
flashing.
Confirm input by
.
Use
to select