16 Instant help
In case of an error in the program process, briefly turn off the balance and discon-
nect from power supply. The weighing process must then be restarted from the be-
ginning.
Fault
Possible cause
The displayed weight does not glow.
• The balance is not switched on.
• The mains supply connection has been
interrupted (mains cable not plugged
in/faulty).
• Power supply interrupted.
• Batteries are inserted incorrectly or empty
• No batteries inserted.
The displayed weight is permanently
changing
• Draught/air movement
• Table/floor vibrations
• Weighing plate has contact with other
objects.
• Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
The weighing result is obviously in-
correct
• The display of the balance is not at zero
• Adjustment is no longer correct.
• Great fluctuations in temperature.
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.