3 adjustment of models cxb_m – KERN & SOHN CXP Version 1.7 User Manual
Page 23

7.6.3 Adjustment of models CXB_M
Operation
Ensure that there are no objects on the weighing plate.
Press
+ 1 + 5 keys
Press
key
The display shows:
0.000
M+
0
12345
Press
key
The display shows:
15.000
(Example)
M+
0
12345
Weight value of adjustment weight is flashing on display.
Carefully place adjusting weight in the centre of the weighing plate.
Press
key
Adjustment will be carried out automatically after dead stop control.
Remove adjustment weight during count down to zero.
The balance returns automatically into weighing mode. An error message will ap-
pear on the display should a adjustment error occur or should the adjustment
weight be incorrect. Turn balance off, then restart it and repeat the adjustment
process.
* The adjustment should be made with the recommended adjustment weight (see
chap. 1 "Technical data"). Weights of different nominal values may be used for ad-
justment but are not optimal for technical measuring.
Info about adjustment weights can be found on the Internet at:
sohn.com
CXB/CXP-BA-e-1417
23