13 error messages – KERN & SOHN DLB 160-3A User Manual
Page 56

56
DLB_A-BA-e-1511
Material
Weight
Sample
(g)
Drying
temperature
( °C)
Drying
period
(approx.)
(min)
Moisture
%
(approx.)
Solid
body
%
(approx.)
Tobacco
1,5
100
16
10,18
Tea, black
2
105
4
7,67
Pasta
1,5
120
8
10,64
Textile fibre
0,8-1,2
85
3,6
14,03
Theophylline
1,5
130
1,9
7,33
Thermoplastic PUR –
granulate
15-18
80
18
0,08
Walnut
2,8
100
5,6
3,5
Washing powder
2
160
12
7,32
Wheat spring water
2-3
90
10
6
Sausage casing
0,2
150
3,5
78,56
Toothpaste
2
100
7,7
34,28
Pulp
2,5
130
4,5
7,32
Cement
8-12
138
4-5
0,8
Sugar
4-5
138
10
11,9
Sugar beets
2
130
13,4
30,94
13 Error messages
ERR01
Weight value instable or zeroing not possible.
Check the environmental conditions.
ERR02
Adjustment error, e.g. instable environmental conditions
ERR03
Adjustment error e.g. incorrect adjustment weight}
ERR05
Data transfer not possible, as weighing value is instable. Check
the environmental conditions.
ERR07
Faulty data reading
ERR10
Instable display when drying starts, check environmental
conditions
ERR11
Sample weight too small
“UNLOAD”:
Sample or sample dish wrongly positioned.
“Err thb”
Heating top does not work, check current supply
Weighing range exceeded, placed load exceeds the capacity of
the appliance. Unload appliance.
Weighing range not reached, e.g. Dish holder / removal help is
missing.