2 adjustment of temperature value – KERN & SOHN DLB 160-3A User Manual
Page 22
![background image](/manuals/697261/22/background.png)
22
DLB_A-BA-e-1511
8.2.2 Adjustment of temperature value
The temperature is measured at two selectable points and it is possible to correct it
there.
Invoke service function „Tmp Adj“, see chap. 8.2
Confirm by pressing the PRINT button. The current setting for the first
temperature point is displayed. Or leave it on factory setting 80°C or select a
desired temperature with the help of the navigation buttons
(available 50-
130°C).
Confirm with the PRINT button, the first heat-up phase is started. The remaining
time is displayed.
Temperature calibration for the first point takes 15 min. Compare the temperature
displayed on the DLB-A01N with that of the moisture analyzer. If the two values
do not match, correct them using the navigation buttons
and confirm with the
PRINT button
.
The current setting for the second temperature point is displayed. Or leave it on
factory setting 150°C or select the desired temperature with the help of the
navigation buttons
. The second temperature point has to be at least 30°C
over the first one, max. 160°C.
Confirm with the PRINT button, the second heat-up phase is started. The
remaining time is displayed.
Temperature calibration for the second point takes 15 min. Compare the
temperature displayed on the DLB-A01N with that of the moisture analyzer. If the
two values do not match, correct with the help of the navigation buttons
.
Confirm by pressing the PRINT button. The adjustment is completed, the
appliance returns into the menu. Use the ON/OFF button to return into
weighing mode.