When an optional printer is connected, the measurement log will be edited
independently on the settings in the menu, see chap. 11.3
For further measurement press the ON/OFF button, the appliance returns into
the menu.
For exiting the menu press anew the ON/OFF button, the appliance returns into
weighing mode
Open the sample chamber and remove the sample with the help of the removal
aid.
Caution: Caution! Sample dish and all parts of the sample chamber are hot!
10.1.3 Moisture analysis with drying program PrG Auto Mode
Drying will be completed, when the set loss of weight (available
0.1 - 9.9% humidity) per time unit (60 sec) is less than the nominal value.
Confirm by pressing the PRINT button. The currently set nominal value is
displayed.
Use the navigation buttons to select the desired setting.
Confirm by pressing the PRINT button. The currently set drying temperature is
displayed.
Use the navigation buttons to select the desired setting.
Confirm by pressing the PRINT button. The query whether the preheating stage
„PrH“ shall be connected, will appear.
Use the navigation buttons to select the desired setting.
PrH no = preheating stage switched off
PrH yes = preheating stage connected