beautypg.com

Carrier 50TCQA User Manual

Page 35

background image

35

downward, into the return air stream. The sampling tube is

connected via tubing to the return air sensor that is

mounted on a bracket high on the partition between return

filter and controller location. (This sensor is shipped in a

flat--mounting location. Installation requires that this

sensor be relocated to its operating location and the tubing

to the sampling tube be connected. See “Completing

Installation of Return Air Smoke Sensor:” for details.)

Return Air Detector Sampling Tube

Controller module

Return Air Detector module
(shipping position shown)*

*RA detector must be moved from shipping position to operating position by installer

C07307

Fig. 61 -- Typical Return Air Detector Location

Return Air with Economizer — The sampling tube is

inserted through the side plates of the economizer

housing, placing it across the return air opening on the

unit basepan. See Fig. 61. The holes in the sampling tube

face downward, into the return air stream. The sampling

tube is connected via tubing to the return air sensor that is

mounted on a bracket high on the partition between return

filter and controller location. (This sensor is shipped in a

flat--mounting location. Installation requires that this

sensor be relocated to its operating location and the tubing

to the sampling tube be connected. See the following

installation procedure.)

Return Air
Sampling Tube

C08129

Fig. 62 -- Return Air Sampling Tube Location

Completing Installation of Return Air Smoke

Sensor:

1. Unscrew the two screws holding the Return Air

Sensor detector plate. See Fig. 63. Save the screws.

2. Remove the Return Air Sensor and its detector plate.
3. Rotate the detector plate so the sensor is facing out-

wards and the sampling tube connection is on the bot-

tom. See Fig. 64.

4. Screw the sensor and detector plate into its operating

position using screws from Step 1. Make sure the

sampling tube connection is on the bottom and the ex-

haust tube is on the top. See Fig. 64.

5. Connect the flexible tube on the sampling inlet to the

sampling tube on the basepan.

Flexible

Exhaust Tubes

Screws

Sample Tube

C08126

Fig. 63 -- Return Air Detector Shipping Position

C08127

Fig. 64 -- Return Air Sensor Operating Position

Additional Application Data — Refer to Catalog No.

HKRNKA--1XA for discussions on additional control

features of these smoke detectors including multiple unit

coordination.

50TC

Q

A