Troubleshooting, 0 troubleshooting – Grant Instruments LTC4 User Manual
Page 36

31115 V1
LTC4
Page 35
Operating Manual
www.grantinstruments.com
11.0 Troubleshooting
Symptom
Possible cause
Action required
Unit fails to cool Compressor
overheated causing
internal over-
temperature
thermostat to operate
Wait for compressor to cool, when thermostat
will reset and compressor will switch on again.
Unit failing to
cool below 5°C
Freezing protection
switch is on and is
preventing further
cooling
Switch off freezing protection (see section
5.2.7).
Unit not cooling
at higher
temperatures
Over temperature
protection switch has
tripped
Refrigeration unit may have temperatures in
excess of 100°C. Leave the unit to cool and
reset (see section 5.2.7)
Cooling
performance
reduced
Condenser fins
clogged
Carry out routine cleaning (see section 9.2)
Temp continues
to rise when not
expected
Set temp is higher
than liquid temp
Check that the bath set temperature is correct
(see section 5.2.10).
Set temperature
too restricted
Liquid type set does
not allow required set
point
Change to different liquid type (see section
5.3.2).
Temperature
does not rise
when expected
Set temp is lower
than liquid temp
Check that the bath set temperature is correct
(see section 5.2.10).
Display shows
“High
Temperature
Alarm”
High temperature
warning alarm has
tripped
Check that the bath set temperature is correct
(see section 5.2.10).
Check that high temperature alarm is correct
(section 5.3.4).
Check that the liquid level in the bath is
adequate (see section 5.1.1 for minimum fill
levels).