LDU 68.1 & 68.2 Technical Manual, Rev. 12 January 2010

Page 7 of 28

4.

Communications and Getting started

4.1. Serial

Interface

Communicating with the LDU 68.1 digitizer is carried out via the RS422/RS485 port.
The data format is the familiar 8/N/1 structure (8 data bits, no parity, 1 stop bit).
Available baud rates via the RS422/RS485 port are as follows: 9 600, 19 200, 38 400, 57 600 or 115 200 baud.

RS422:

Connection using a 4 wire technique

Point-to-Point connection, i.e. no bus communication possible

Half duplex setup (DX=0)

RS485:

Connection using 2- or 4-wire techniques

Multi-drop connection possible, up to 32 LDU XX.X

Half or Full Duplex (DX=0 or DX=1) possible

(RS232):

The optional adapter board UA 77.1 is available which has an RS485/RS232 converter built in.

4.2. Command

Language

The command set of LDU XX.X series is based on a simple ASCII format (2 capital letters). This enables the
user to setup the device, get results or check parameters.

Example: LDU XX.X with the address or channel number 1 is connected via the RS 485 port to a bus system.
You want to get the net weight.

In this manual means: Space “_“ and Enter (CR/LF) ““

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

OP 1

Open device #1

OK

Device # 1 ready

GN

Get net weight

N+123.45

Net weight with algebraic sign / floating point

The command OP_2 opens the communication channel to LDU XX.X device #2. Now device #2 acknowledges
that it is active and responds to any commands on the bus. Communication with device #2 will be closed by
another OP command (for another device on the bus e.g. OP_5) or by the command CL_2.

Each OP_X command implies a CL command to all other devices on the bus except device #X. This makes the
address structures easier and the system performance improves.