8.7. Auto–transmit

Commands – SG, SN, SF, SW, SA

The following commands allow the gross weight or net weight values to be continuously sent. Continuous
transmission start as soon as the relevant command has been issued and finishes when any other valid
command is accepted by the LDU 68.1. The data output rate will depend on the baud rate being used e.g. with
a baud rate of 9 600, approximately 100 values per second can be transmitted.

Note:

The SG, SN and SW commands will only work if the LDU 68.1 has been set to full duplex [DX=1].

The continuous transmission of either the gross or net values will stop when another valid command is
received.

8.7.1. SG Send Gross Value continuously

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

SG

G+01.100

Gross value: 1,100 d

8.7.2. SN

Send Net Value continuously

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

SN

N+01.000

Net value: 1,000 d

8.7.3. SF

Send Filtered Value continuously (Display Mode only)

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

SF

F+01.000

Filterted net value: 1,000 d

This command will start to auto-transmit the post-filtered net weight value, e.g. for a weight display. The filter
setup will be done with the command FF (see chapter 8.4.).

8.7.4. SW Send Data String “Net, Gross and Status“ continuously

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

SW

W+000100+0011005109

(example)

Net value: +000100 d (no decimal point)

Gross value: +001100 d (no decimal point)

Status bit 1: 5 (not used)

Status bit 2: 1 (Hex)

Check sum: 09 (Hex)

The status bits 1 and 2 are defined as follows:

Status

Value = 1

Value = 2

Value = 4

Value = 8

Status bit 1 Not used

Not used

Output 0 active

Output 1 active

Status bit 2 Signal stable

Set zero performed

Tare active

Not used

The check sum is the reciprocal value of the sum of all ASCII values within the data string without the check
sum itself.

8.7.5. SA

Send Triggered Average Value automatically

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

SA

A+01.100

Measurement value: 1,100 d

This command will start to auto-transmit the measurement value of the current trigger cycle. The trigger setup
commands are described in chapter 8.12.