8.4. Filter Setting Commands – FL, FF, UR

A digital filter can be set which will eliminate most of the unwanted disturbances. The commands FL and FF are
used to define the digital filter settings, the command UR is used to define an averaging of up to 90
measurement values. Please note that these filters are positioned immediately after the A/D Converter and
therefore affect all aspects of the weighing operation.

8.4.1. FL

Filter

Settings

This command defines the filter cut off frequency and the filter characteristics.

The digital IIR filter operates as

2

nd

order low pass filter and Bessel characteristics. The attentuation is 40dB/decade (12 dB/octave).

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

FL

F+00003

Request: FL = 3

FL 1

OK

Setup: FL = 1

WP

OK

Setup saved

The permitted settings are from 0 to 7 (see table below

).

Factory default setting: FL = 3 (Bessel 0.5 Hz).

Filter Cut-off Frequencies
FL

=

7 6 5 4 3 2 1 0

Cut-off frequency

0.02 Hz

0.05 Hz

0.1 Hz

0.2 Hz

0.5 Hz

1 Hz

2 Hz

5 Hz

8.4.2. FF

Moving

Average

This command defines a moving average for the output value. The permitted settings are from 0 to 15 (see
table below

). Factory default setting: FF = 0 (0.2 s)

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

FF

F+00001

Request: FF = 1 (0.4 s)

FF 4

OK

Setup: FF = 4 (1 s)

WP

OK

Setup saved

Moving Average
FF

=

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

sec

0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2

8.4.3. UR Update

Rate

This command defines the update rate of the output value. The permitted settings are from 0 to 2 (see table
below

). Factory default setting: UR = 2 (30 values/s)

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

UR

U+0000

Request: 90 values/s active

UR 2

OK

Setup: 30 values/s

WP

OK

Setup saved

Update Rates

UR

0 1 2

Values / second

90

45

30