LDU 68.1 & 68.2 Technical Manual, Rev. 12 January 2010

Page 24 of 28

8.11. Save Calibration and Setup Data Commands – CS, WP, SS, GI, PI

The calibration and setup parameters can be divided in 3 groups:

Calibration

: CMn, CI, MR, DS, DP, CZ, CG, ZT, ZR, FD, LC and LNn, saved by command CS

Setup

: FL, FM, NR, NT, BR, AD, DX and other, saved by command WP

Setpoints:

S0, S1, H0, H1, A0, A1, saved by command SS

Note:

Calibration data can only be saved if the TAC code is known and precedes the CS command. See the

CE

and CS commands in chapter 8.2.

The setup data and the setpoint data will be stored non-volatile in the EEPROM using the WP respective SS
command.

8.11.1. WP Save the Setup Parameters

With this command the settings of the “Filter” (FL, FM) ,the “No-motion” (NR, NT) and the communication (AD,
BR, DX) will saved in the EEPROM.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

WP

OK

Setup data saved

WP

ERR

Error

8.11.2. SS

Schaltpunkt-Parameter sichern

With this command the setpoints (S0, S1), the setpoint hysteresis (H0, H1) and the setpoint allocation (A0, A1)
will be saved in the EEPROM.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

SS

OK

Setpoint parameters saved

SS

ERR

Error

8.11.3. GI

Save an Image File from the EEPROM

Saves a HEX-INTEL formatted EEPROM image file from the source LDU’s EEPROM. The image file contains
all stored information except the calibration data. This image file can be downloaded to any LDU with the same
firmware type and revision no. as the source LDU.

8.11.4. PI

Download an Image File to the EEPROM

Downloads a HEX-INTEL formatted EEPROM image file to the target LDU’s EEPROM. The image file contains
all stored information except the calibration data.

Attention:

The target LDU must have same firmware type and revision no. as the source LDU.