Calibration and Calibration Sequence

The calibration of LDU 68.x is only possible after starting a calibration sequence (compare with chapter 8.2).

Command CE:

Calibration enable – returns the current TAC value

Command CM:

Calibrate maximum display – sets the max. allowable display value

Command CI:

Calibrate minimum – sets the minimum allowable display value

Command DS:

Display step size – sets the output incremental step size

Command DP:

Decimal point – sets the position of the output decimal point

Command CZ:

Calibrate zero – sets the system zero point

Command CG:

Calibrate gain – sets the system gain

Command ZT:

Zero track enable / disable

Command FD :

If applicable: Factory default setting (return to)

Command

CS

:

Calibration

save

(TAC

counter automatically incremented by 1)



Preparing the calibration:

Check, if the max value of the display is set sufficiently high (see chapter 8.2: command CM)

Check, if the no motion conditions are defined reasonable (chapter 8.3: e.g. NR = 1, NT = 1000)

Set the filter frequency to 0.5 Hz (see chapter 8.4: FL = 3)



Example:

Setup of zero point, system gain and decimal point

The chosen calibration weight has the value 5000 (increments). That could be 500 g, 5 kg or 5000 kg. We
calibrate with 500 g. The decimal point is set up by command DPx (x = 1, 2 or 3), here 1 figure after the
decimal point. A measured weight of 500 g is displayed as 500.0.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

CE

E+00017

(example)

Request: TAC counter CE17

Adjust zero: The scale has to be empty. No load!

CE 17

OK

Calibration sequence active

CZ

OK

System zero point saved

Adjust gain: First put the calibration weight on the scale (here 500 g)!

CE 17

OK

Calibration sequence active

CG 5000

OK

Setting span

CG

G+05000

Request: span 5000 d

CE 17

OK

Calibration sequence active

DP 1

OK

Setting: decimal point 0000.0

CE 17

OK

Calibration sequence active

CS

OK

Save calibration data in EEPROM


Zero point, gain and decimal point position were saved in the EEPROM; the calibration counter (TAC) is
increased automatically by 1.