HI HLPT SERIES LOAD POINT ASSEMBLY

Page 10

Step 2.

Use a megohmmeter and measure the

resistance between all five wires tied

together and the load cell metal body.

The measured value should be

5,000 megohms or more.

WARNING

WHEN USING A MEGGER DO NOT EXCEED
50 VOLT RANGE.

If the sensor fails this test

remove the ground wire and

test with only the four live

leads.

If the sensor passes the test an

insulation problem in the cable

is most likely.

Step 3.

Replace the load cell if the cell fails both

tests.

Electrical Termination
Cable Color Codes

The cable is 6 conductor, shielded (floating) and 20

feet in length.

EXC+

Red

EXC -

Black

SIG +

Green

SIG -

White

C2+

Gray

C2-

Violet

SHIELD

Yellow

WARNING

LOAD CELL CABLE LENGTH HAS BEEN
CALCULATED INTO C2 CALIBRATION
DATA. HARDY PROCESS SOLUTIONS REC-
OMMENDS THAT YOU DO NOT CUT YOUR
ADVANTAGE OR ADVANTAGE LITE LOAD
SENSOR CABLE, AS YOUR C2 ACCURACY
WILL BE AFFECTED AND THE WARRANTY
WILL BE VOIDED.