ChemoMetec SCC-400 User Manual

Page 33

background image

6!NucleoView!SCC0400!!

19!

6.2.3 Tools-L-Calibration-
The!NucleoCounter!SCC0400!system!is!carefully!calibrated!to!give!stable!cell!counting!
results!and!all!SCC0400!instrument!will!give!similar!results!when!measuring!the!same!
sample.!If,!on!the!other!hand,!the!results!need!to!be!adjusted!then!the!Tools%U%Calibration!
menu!item!can!be!used!to!adjust!the!respons.!!

!

Figure,21,.,Tools,.,Calibration,menu,item,

The!Calibration!function!opens!the!dialogue!shown!below.!Here!it!is!possible!to!define!
Slope!and!Intercept,!which!are!applied!to!the!measurements.!The!calibrated!results!(SCC')!
will!be!as!follows:!!
!

!

!

!

SCC'!=!Intercept!+!SCC

*

Slope!

!

Figure,22,.,The,Calibration,dialogue,

6.2.4 Tools-L-Instrument-
There!are!some!aspects!of!the!operation!of!NucleoCounter!SCC0400!that!can!be!defined!
using!the!Tools%U%Instrument!menu!item.!

!

Figure,23,.,Tools,.,Instrument,menu,item,

The!Instrument!function!opens!a!dialogue!as!shown!in!the!following!figure.!In!the!dialogue!
it!is!possible!define!the!precision!mode,!priming!time!and!thickness!of!sample!
compartment.!