Step 4b: pull-off test – Checkline AT-A User Manual
Page 7
Page 8
Dolly Size
PSI Rates
MPa Rates
10 mm
125, 200, 400, 600, 725
1.00, 2.00, 3.00, 4.00, 5.00
14 mm
60, 100, 200, 300, 360
0.40, 0.70, 1.40, 2.00, 2.50
20 mm
30, 50, 100, 150, 180
0.20, 0.30, 0.70, 1.00, 1.20
50 mm
5, 8, 16, 24, 30
0.04, 0.08, 0.12, 0.16, 0.20
Quick Guide
(1) Connect the actuator to the dolly
(2) Power-up with the button
(3) Verify measurement units, dolly size and pull rate
(4) Ready the instrument with the button
(5) Test with the button
1. Place the actuator assembly over the dolly head and attach the quick coupling
to the dolly by reaching through the holes in the actuator assembly and lifting the
quick coupling. Release the quick coupling when the dolly head is completely
engaged.
2. Press the button to power-up the instrument if necessary. The instrument
will power-down after 5 minutes of no activity or by holding the button for 2
seconds.
3. Check settings:
Step 4b: Pull-Off Test
Step 4b: Pull-Off Test
(PosiTest AT-A Automatic
)
LCD
USB port
Actuator
Handle
Quick
Coupling
Actuator
Assembly
Hose
AC Adapter Port
(Charges batteries
and powers instrument)
(a) Verify displayed measurement units. Change with the button if
necessary.
(b) Verify dolly size with the button and change if necessary.
(c) Verify pull rate with the button and change if necessary. The following
user selectable rates are available: