Checkline LM6 User Manual

Page 11

background image

– 10 –

4. Press the third button as long as the black bar stands

on the first digit of the serial number. After a waiting

time of a few seconds the device accepts the entered

digit. Now the other 3 digits have to be entered.

When the 4-digit number has been entered correctly

(see picture), press the

button to confirm.

5. The request Reinitialize? will appear at the display.

Hold the humimeter LM6 in one hand and make sure

that nothing stands behind the measuring field (black

plate).

6. Now press the second button (

) with one finger. The

calibration procedure will take a few seconds. During

that time, the instrument has to be held up in the air

and the measuring field must NOT be touched.

7. The device is ready when the shown picture on the

right disappears automatically.