2 system menu – Checkline AWS-4050 User Manual
Page 9
AWS-4050 Guide
1-2 Filter
+, - Set filter value in Hz.
125, 250, 500, 1000, 1500 & 2000.
1-3 Peak Blanking
+, - Set blanking in percent of full scale.
Peak Blanking sets the minimum threshold at which torque is captured
as a peak. Pressing the + and - buttons will toggle this from 2% through 50%
of Full Scale. Pressing the ENT button will accept this value and return to Run
Mode.
1-4 Sign Lock
+, - Toggle On or Off.
The Sign Lock feature allows the user to select the torque direction of
the peak to be captured. With Sign Lock ON, the initial direction or sign is the
default direction. Any peak measured in the non-selected direction will be
measured, but not captured as a peak. To reset the sign, either cycle the
power or zero the unit.
With Sign Lock OFF, the tester will capture both CW and CCW torque.
1-5 Edit Limits
High or Low edits that limit.
Limit editing:
Up and Down: Change the digit specified (5 is decimal point, OFF turns
the limit off)
DIG: Changes which digit to edit.
ENT: Accepts changes.
2 SYSTEM MENU
SEL: Enters menu.
2-1 Sleep
+, -: Set time in minutes.
To conserve battery life, the display is equipped with a "sleep" mode,
which sets the limit to standby after a user-settable amount of time has
passed without activity. Press the + key to increase the number of minutes the
unit will wait for input, or the - key to decrease the number. The highest