Checkline AWS-4050 User Manual
Page 12
AWS-4050 Guide
the eight possible choices: Kgf m, KgfCm, gfCm, cNm, Nm, FT LB, IN LB, IN
OZ.
F
ULL
S
CALE
This screen shows the Full-Scale value of the Torque Shaft. This is not a field
adjustable value.
L
OW
L
IMIT
Use the low limit setting as a means of visually flagging the operator when a
reading fails to reach a desired minimum value. A small down arrow will
appear on the screen if a peak is captured below the limit setting.
The limit is adjusted by using the navigation buttons to set the first four digits
to the desired value.
The fifth digit is used to select the decimal point
position. The up and down buttons under the “Soft” button Change will
change the value from 0-9. Pressing the “soft” DIG will scroll through the digit
positions.
When the correct value is entered, pressing the ENT button will
return you to the menu system.
Once all the digits have been set, press the MEM key to accept the value and
return to Program Function mode. The next time a reading is taken, "LO" will
appear on the display if the captured value is less than the low limit
H
IGH
L
IMIT
Use the high limit setting as a means of visually flagging the operator when a
reading falls over a desired maximum value. High limits are set in the identical
way as low limits. Please refer to the Low Limit section for details.
NOTE ON LIMITS: The green LED on the front of the display will flash when a
peak is captured that falls within the limit setting.