Description of functions – Checkline AWS-4050 User Manual

Page 11

background image

AWS-4050 Guide

3-3 Reset All

CLR: Followed by ENT clears all readings and all sets. Should be used

approximately every 5000 readings.

DESCRIPTION OF FUNCTIONS

The following is a description of the standard features of the AWS product line.

O

PERATING

M

ODE

Current operating mode (Peak, 1st Peak or Track) will show on the display.
Press the Œ key to toggles between them.

Peak

Displays and retains the maximum torque experienced by the wrench, as

occurs when operating the wrench in the tightening direction. The Peak Mode

is used for all power tools and some dial wrenches.

1st Peak

Detects the “first peak” of torque experienced by the wrench, capturing the
initial torque as occurs when the torque wrench cams over. First Peak is used

primarily for Click torque wrenches and cam over screwdrivers.

Track

Displays torque as it is being applied to the transducer. Track mode is used

primarily for verifying calibration of the unit.

S

ELECTING A

T

RANSDUCER

IMPORTANT NOTE: On power-up, the transducer that is selected will be active

and ready to use. When changing the shaft, you MUST Zero the unit. Do this

by pressing the On/Zero. The shaft capacity can be verified by going to the

Information screen in the Systems Menu (2-4). Full Scale capacity screen is

shown. Press the ENT button twice to return to the run mode.

E

NGINEERING

U

NITS

Shows

the

current

engineering

units.

Press the key to cycle through