TORQUE MEASUREMENT

page 4

TORQUE MEASUREMENT

Output is available in PD, PP and C modes. In PD and PP modes peak
data is output when ZERO is pressed or activated by the AUTO ZERO
function. In C mode, the gauge outputs data continuously 180
data/second.

GENERAL OPERATION

IMPORTANT! Always make sure that the sensor is mounted securely
to a stable workbench to avoid spinning and possible accident.

1. Connect the torque sensor to the tester dis-

play unit with cable provided. When con-
necting the cable to the torque sensor, rotate
the round connector to find the matching
groove, then push in until it makes positive
connection (When disconnecting the cable
from the torque sensor, hold the silver metal
part of the round connector and pull out).

2. To select the desired measuring unit (lbf-in,

kgf-cm, N-m, N-cm), press POWER, then
hold UNITS for 4 four seconds until a beep sounds. Press UNITS
to cycle choices. Press PRG/SET to enter.

3. Press and hold MODE for one second to select from the follow-

ing Measuring Modes.
Real Time–Displays torque transients (no output or indicator)
PP Mode– Peak, captures peak torque (peak data output, PP

appears on display)

PD Mode– First Peak, capture first peak value (peak data

output, PD appears on display)

C Mode–

Continuous RS-232 output Display and output
torque transients (180 data/sec., GOOD indicator

flashes)

4. Insert wrench drive securely into the sensor, hold perpendicular-

ly (not at an angle), and turn to measure.

5. After measuring, press CLEAR to zero display for the next test.