TORQUE MEASUREMENT

page 8

TORQUE MEASUREMENT

Auto Power Off

To maximize the life of the battery, power will automatically shut
off after 10 minutes of non-use.

System Reset Button

When battery has been depleted completely and a recharge has
been executed, the system may not yet work. In this case, press the
System Reset Button

.

Low Battery Indicator

“–L– –” will appear when tester needs to be recharged. Turn off
power.
IMPORTANT! Use the provided IMADA/CEDAR AC adapter/charger
exclusively and plug into the correct AC output. It takes 8 hours to
charge fully. Do not recharge for more than 12 hours. When it is
fully charged, make sure the AC adapter/charger is disconnected to
avoid overcharging.

In order to avoid high heat, explosion or toxic fumes, please note
the following precautions:

Be sure to use the provided AC charger exclusively.
Do not plug the charger into unspecified higher voltage.
Do not recharge until LOBAT icon appears.

8

Auto Zero Reset Switch

for Peak or Peak Down mode

Rather than resetting manually for each test, the Automatic Zero
Reset can be programmed by using a paperclip to push the Auto
Zero Reset Switch

.

As the Auto Zero Reset Switch

is pressed (use a pointed tool,

like a paperclip end), the Automatic Zero Reset durations cycle.
(0.0C

Þ 0.5C Þ 1.0C Þ 1.5C Þ 2.0C Þ 2.5C Þ 3.0C) Select the

desired duration. For example, 0.5C means 0.5 second and 0.0C
means MANUAL RESET. If a selection is not made within 3 seconds,
the cycle begins again.

12

12

Virtual RS-232C Serial Output

Output is available in PD, PP and C modes. In PD and PP modes peak
data is output when ZERO is pressed or activated by the AUTO ZERO
function.

Comport signal

8 data, 2 stop, no parity. Baud Rate: 19,200 bps.

Output Connector

USB A/B cable

Peak Data Output Format

[CAN] _ _ _ [SO] [value] _ [SI] [unit] [CR]

(_ _ _ is memory location)

[CAN]:

ASCII control code 24

_:

Space (code 32)

[SO]:

ASCII control code 14

[value]:

Output data with sign and decimal point. Plus sign
represents for CW torque and minus sign for CCW.
[Value] always occupy six locations and empty
locations will be filled with spaces.

[SI]:

ASCII control code 15

[unit]:

N*m _ _ _=N.m
kgf*cm = kg.cm
lb*in _ = lb.in

[CR]:

ASCII control code 13 (Carriage Return)

Continuous Output Data Format

[CAN] [value] [CR]