TORQUE MEASUREMENT
page 8
TORQUE MEASUREMENT
Auto Power Off
To maximize the life of the battery, power will automatically shut
off after 10 minutes of non-use.
System Reset Button
When battery has been depleted completely and a recharge has
been executed, the system may not yet work. In this case, press the
System Reset Button
.
Low Battery Indicator
“–L– –” will appear when tester needs to be recharged. Turn off
power.
IMPORTANT! Use the provided IMADA/CEDAR AC adapter/charger
exclusively and plug into the correct AC output. It takes 8 hours to
charge fully. Do not recharge for more than 12 hours. When it is
fully charged, make sure the AC adapter/charger is disconnected to
avoid overcharging.
In order to avoid high heat, explosion or toxic fumes, please note
the following precautions:
Be sure to use the provided AC charger exclusively.
Do not plug the charger into unspecified higher voltage.
Do not recharge until LOBAT icon appears.
8
Auto Zero Reset Switch
for Peak or Peak Down mode
Rather than resetting manually for each test, the Automatic Zero
Reset can be programmed by using a paperclip to push the Auto
Zero Reset Switch
.
As the Auto Zero Reset Switch
is pressed (use a pointed tool,
like a paperclip end), the Automatic Zero Reset durations cycle.
(0.0C
Þ 0.5C Þ 1.0C Þ 1.5C Þ 2.0C Þ 2.5C Þ 3.0C) Select the
desired duration. For example, 0.5C means 0.5 second and 0.0C
means MANUAL RESET. If a selection is not made within 3 seconds,
the cycle begins again.
12
12
Virtual RS-232C Serial Output
Output is available in PD, PP and C modes. In PD and PP modes peak
data is output when ZERO is pressed or activated by the AUTO ZERO
function.
Comport signal
8 data, 2 stop, no parity. Baud Rate: 19,200 bps.
Output Connector
USB A/B cable
Peak Data Output Format
[CAN] _ _ _ [SO] [value] _ [SI] [unit] [CR]
(_ _ _ is memory location)
[CAN]:
ASCII control code 24
_:
Space (code 32)
[SO]:
ASCII control code 14
[value]:
Output data with sign and decimal point. Plus sign
represents for CW torque and minus sign for CCW.
[Value] always occupy six locations and empty
locations will be filled with spaces.
[SI]:
ASCII control code 15
[unit]:
N*m _ _ _=N.m
kgf*cm = kg.cm
lb*in _ = lb.in
[CR]:
ASCII control code 13 (Carriage Return)
Continuous Output Data Format
[CAN] [value] [CR]