page 10

TORQUE MEASUREMENT

TORQUE MEASUREMENT

ON/OFF Switch Press to turn on, press again to turn off (click
once, do not hold). After 10 minutes of non-use the unit shuts off.

Mode Select Switch
PD (Peak Down):
To capture critical first peak value.
C (Real Time): To display torque transients.
PP (Peak): To capture peak torque achieved during a test.

CLEAR Switch To reset the display or output peak data

Zero Adjustment Pot.

LED Display Displays torque and low battery icon (LOBAT).

Torque Value

Number of Blows

Sensor Cable Receptacle

AC Charger/Adapter Receptacle If LOBAT icon appears,
8-hour battery recharge is required.

System Reset Button When battery has been depleted com-
pletely and a recharge has been executed, the system may not
yet work. In this case, press the System Reset Button.
Unit Switch To select measuring units (lbf-in, kgf-cm & N-cm)

USB port (Virtual RS-232 Output)

Blow Counter Duration Timer Use a paper clip to push the
switches, right to increase, left to decrease.

Auto Zero Reset Switch

1

2

3

4

5a

5b

6

7

8

9

10

12

11

1

2

3

4

5a

5

5b

6

7

8

9

10

11

12

DI-1M-IP SPECIFICATIONS

DI-1M-IP50

DI-1M-IP200

DI-1M-IP500

Measuring Range

0.30 50 N-m

3.0 200 N-m

3.0 500 N-m

3.0 500 kgf-cm

30 2000 kgf-cm

30 5000 kgf-cm

3.0 430 lbf-in

30 1800 lbf-in

30 4500 lbf-in

Peak Down

0.70N-m

7.0N-m

Operating Range

7.0kgf-cm/lbf-in

70kgf-cm/lbf-in

Accuracy

±0.5 F.S.

Display

4 digit LED

Operation

Both CW and CCW

Measuring Modes

P-P Peak Mode displays PEAK torque value
that will not change until a higher value is measured.
T-R Real Time Mode displays TRANSIENT torque values.
P-D Output Torque Data every 1/87 seconds.

Auto Clear

Automatically zero the display, 0.5 - 3.0 sec,
selectable at 0.5 sec intervals (0.0 sec for manual)

Blow Measurement

0 - 99 blows; 0.1 - 9.9 seconds

Data Output

ASCII format (baud rate 19,200)

Power Supply

Ni-Cad rechargeable battery 1.2Vx5cells (700mAh)

Auto Shut Off

Automatic shut off after 5 minutes of non-use

Charging Time

8 hours

Working Time

8 hours continuous

Meter Weight

Approx. 600g

Sensor Weight

Approx. 600g

Approx. 1kg

Approx. 2.5kg

Meter Dimensions

146mm(W) x 115mm(D) x 48mm(H)

Accessories

Carrying case, AC/DC adapter