OPERATION 5

RS-232

RS-232

The AWS-QC display can be connected to a printer, computer or data
collector via its RS232 interface. Every time a reading is accepted into
memory, a peak is cleared, or data is transmitted via the print data
menu(s), it is transmitted via the RS-232 port. To download the
readings, go to the DATA MENU. Cables are available directly from AWS,
or they can be constructed according to the following tables. AWS also
offers various software packages to get your readings into excel or other
programs. Call us for more information

The AWS-QC display can be connected to a printer, computer or data
collector via its RS232 interface. Every time a reading is accepted into
memory, a peak is cleared, or data is transmitted via the print data
menu(s), it is transmitted via the RS-232 port. To download the
readings, go to the DATA MENU. Cables are available directly from AWS,
or they can be constructed according to the following tables. AWS also
offers various software packages to get your readings into excel or other
programs. Call us for more information

RS232 Transfer Protocol

RS232 Transfer Protocol

Protoco

l

Protoco

l

Value

Value

Cable 9 pin to mini-

plug.

Baud 9600

Parity None

Bits 8

S Bit 1

Flow None

RS232 Datastream Format

mmmbsdddddbuuuuucl,

where:

m Memory Location

s Sign (space or -)

c Carriage Return

d Data with Decimal

Point

l Line Feed

u Units

b Blank

RS232 Cable Pinouts

Pin
#

Description

Pin
#

Description

1 Unused

6 Unused

2

Transmit

7

Unused

3 Receive

8 Unused

4

Unused

9

Unused

5 Ground








AWS-QCM Guide