Rs-232, Rs232 transfer protocol, Rs232 datastream format – Checkline TT-QCM User Manual
Page 6: Rs232 cable pinouts
OPERATION 5
RS-232
RS-232
The AWS-QC display can be connected to a printer, computer or data
collector via its RS232 interface. Every time a reading is accepted into
memory, a peak is cleared, or data is transmitted via the print data
menu(s), it is transmitted via the RS-232 port. To download the
readings, go to the DATA MENU. Cables are available directly from AWS,
or they can be constructed according to the following tables. AWS also
offers various software packages to get your readings into excel or other
programs. Call us for more information
The AWS-QC display can be connected to a printer, computer or data
collector via its RS232 interface. Every time a reading is accepted into
memory, a peak is cleared, or data is transmitted via the print data
menu(s), it is transmitted via the RS-232 port. To download the
readings, go to the DATA MENU. Cables are available directly from AWS,
or they can be constructed according to the following tables. AWS also
offers various software packages to get your readings into excel or other
programs. Call us for more information
RS232 Transfer Protocol
RS232 Transfer Protocol
Protoco
l
Protoco
l
Value
Value
Cable 9 pin to mini-
plug.
Baud 9600
Parity None
Bits 8
S Bit 1
Flow None
RS232 Datastream Format
mmmbsdddddbuuuuucl,
where:
m Memory Location
s Sign (space or -)
c Carriage Return
d Data with Decimal
Point
l Line Feed
u Units
b Blank
RS232 Cable Pinouts
Pin
#
Description
Pin
#
Description
1 Unused
6 Unused
2
Transmit
7
Unused
3 Receive
8 Unused
4
Unused
9
Unused
5 Ground
AWS-QCM Guide