beautypg.com

Description of functions, Operating mode, Peak – Checkline TT-QCM User Manual

Page 14: 1st peak, Track, Engineering units, Full scale, Low limit

background image

Description of Functions 13

DESCRIPTION OF FUNCTIONS


The following is a description of the standard features of the AWS
product line.

O

PERATING

M

ODE

Current operating mode (Peak, 1st Peak or Track) will show on the
display. Press the Πkey to toggles between them.

Peak

Displays and retains the maximum torque experienced by the wrench, as
occurs when operating the wrench in the tightening direction. The Peak
Mode is used for all power tools and some dial wrenches.

1st Peak

Detects the “first peak” of torque experienced by the wrench, capturing
the initial torque as occurs when the torque wrench cams over. First
Peak is used primarily for Click torque wrenches and cam over
screwdrivers.


Track


Displays torque as it is being applied to the transducer. Track mode is
used primarily for verifying calibration of the unit.

E

NGINEERING

U

NITS

Shows the

current engineering units.

Press the key to cycle

through the eight possible choices: Kgf m, KgfCm, gfCm, cNm, Nm, FT LB,
IN LB, IN OZ.


F

ULL

S

CALE

This screen shows the Full-Scale value of the Torque Shaft. This is not a
field adjustable value.

L

OW

L

IMIT

Use the low limit setting as a means of visually flagging the operator
when a reading fails to reach a desired minimum value. A small down
arrow will appear on the screen if a peak is captured below the limit
setting.

AWS-QCM Guide