beautypg.com

Checkline MX-500 User Manual

Page 2

background image

2

7

SPECIFICATIONS
Motorized Vertical Test Stand
Model No:

MX-500

Max Load:

500 lbf

Speed:

0.2 – 7 in/min

Stroke:

13" (330 mm)

Power:

115/230 VAC

FS

MODE

Manual Operation
Single Cycle
Continuous Cycle
FS Effective

SWITCH

SPEED

EMERGENCY

FS Volume

INTRODUCTION
The Model MX-500 Vertical Motorized Test Stand assures
consistent, identical measuring conditions by eliminating
possible human error.

CONTROL PANEL

Power Switch
Adjustable Speed Direction

Switch

Maximum Speed Direction

Switch

Speed Adjust Knob

Emergency Stop Switch
Mode Switch
Mode Indicators
Fixed Speed Switch
Fixed Speed Adjustment

TRAVEL LIMITS

UPPER TRAVEL LIMIT
Cross Head stops moving up when it
reaches this position. To adjust the
stroke, loosen screw and move the limit
to desired position. Re-tighten screw
securely.

MIDDLE TRAVEL KNOB
Slows the cross head to the speed set by
the Speed Adjust Knob.

LOWER TRAVEL LIMIT
Cross Head stops moving down when it
reaches this position. To adjust the
stroke, loosen screw and move the limit
to desired position. Re-tighten screw
securely.

CAUTION:
Make sure to set the lower travel limit high enough to allow
adequate clearance for the force gauge and attachments, so it
will not be overloaded.

To prevent accidental movement of the lower limit position,
replace the thumbscrew with an M5 Allen screw so that an Allen
wrench will be required to adjust it.

CROSS
HEAD

RESET FROM OVERLOAD

If the test stand overloads or locks up during testing, disconnect
the cable connecting the test stand to the force gauge, switch to
Manual Operation and using a direction switch, back off the
crosshead (in the opposite direction from the overload) to
release the load.