Selecting High and Low Setpoints

Program High and Low setpoints to enable GO/NG testing.

1. Turn off the gauge.

2. Hold

and press

once. The LCD display will

briefly show HI, then the High setpoint value with flashing

HI. Press

to increase and

to decrease the High

value, then press

to select. The LCD display briefly

shows Lo and then Low setpoint value with flashing L.

Press to

increase and

to decrease the Low value,

then press

to select.

For example, 5 lbf is set as Low setpoint and 10 lbf as the
High setpoint, the GO/NG indicator on the LCD display
shows

for measurements less than 5 lbf (Low setpoint).

for measurements between 5–10 lbf and

for over

10 lbf (High setpoint).

+

O

OPTIONAL SETTINGS

The following steps are not necessary for most test applications and
may be skipped entirely.

Selecting Test Type

(continued from step 2)

Select rapid data processing for Destructive Testing.

3. After

is pressed in step 2, the display briefly shows

F-AdC

and then FA with flashing F.

4. Press

or

to cycle between FA (for Destructive

Testing) or SL (for Non-destructive Testing: Factory default),

then press

to select.The LCD display briefly shows

the gauge capacity and then zero.

SL

: Non-destructive Testing

(for most testing, 30 data/second)

FA

: Destructive Testing

(for very rapid force changes, 1,000 data/second)