beautypg.com

Mode 5: self test, Error codes, Display switches & functions – Checkline DT-5TG User Manual

Page 7

background image

Mode 5: Self Test

This mode lets you check the LED display and input
circuitry.

1. Press mode and data set keys for

five seconds.

2. Press increment key and select

mode 5.

3. Press mode key to test display

segments and decimal points.

4. Some display functions switches

can also be checked. Press these
keys to test:

MAX

MIN

PEAK

HOLD

5. Press mode key. The display will

show

1000

1000

1000

1000

1000 and signal LED

LED

LED

LED

LED will

flash.

6. Press data set key. The display will

go back to the previous mode of
operation.

Error Codes

Display

Type of Error

What to Do

EE-00

EE-00

EE-00

EE-00

EE-00

Parameter setting

Press data set key. Enter parameter according to setting range.

EE-01

EE-01

EE-01

EE-01

EE-01

Hi/Lo setting

Press data set key. Enter the upper and lower limits.

EE-02

EE-02

EE-02

EE-02

EE-02

Internal setting

Press data set key. Interrupt power at terminals 1 and 2.

EE-03

EE-03

EE-03

EE-03

EE-03

Memory recall

Press data set key. In sequence, press HI, hold, increment and mode keys.

EE-04

EE-04

EE-04

EE-04

EE-04

Communication

Check connections with module. Press data set key.

EE-05

EE-05

EE-05

EE-05

EE-05

error with

EE-06

EE-06

EE-06

EE-06

EE-06

module

Display Switches & Functions

The DT-5TG has seven display functions located above
the display. All functions have an LED indicator and all
but GO have a membrane switch. Here is a brief descrip-
tion of each function:

HI

If the display is equal to or greater than the HI
limit setting, the LED will light and a contact
will close.* To program, press HI and mode keys
for one second. Use the increment and shift
keys to set limit.

GO

If the display is between the HI and LO settings,
the LED will light and a contact will close.*

LO

If the display is equal to or less than the low limit
setting, the LED will light and a contact will
close.* To program, press LO and mode keys for
one second. Use increment and shift keys to set
limit.

MAX

The display will hold the average maximum
measurement. To program, press MAX and
mode keys until LED lights.

MIN

The display will hold the average minimum
measurement. To program, press MIN and
mode keys until LED lights.

PEAK The display will hold the absolute peak measure-

ment. To program, press mode and PEAK keys
until LED lights.

HOLD This function will hold the display indefinitely as

long as the unit is powered. To program, press
mode and HOLD keys until LED lights.

*Requires the DOP-CP Triple Relay Output Module.

Note:

!

The HI, GO and LO functions are non volatile
and may be reset by programming to zero.

!

The MAX, MIN, PEAK and HOLD functions
must be used separately. These functions may be
reset by pressing the data set key or by interrupt-
ing power.

7