8 .3 reference measurement fails – BUCHI NIRMaste User Manual
Page 59

8 Troubleshooting
59
NIRMaster Operation Manual, Version A
8 .3
Reference measurement fails
When the deviation between the stored reference spectrum is out of limit, the reference measurement
will fail. A pop-up window will show both curves and give details of the measurement result.
Possible reasons for a reference measurement to fail
Optical surfaces (e.g. External reference) dirty or scratched
•
Environmental conditions out of operational range (e.g. condensating humidity at optical surfaces)
•
Defective instrument (e.g. electronic damage)
•
After lamp-module exchange
•
Remedy
Clean the External reference and the NIRMaster output window
•
Place the External reference onto the output window and press the ‘Rerun’ button to repeat the
•
measurement
If measurement still fails, a new reference spectrum can be stored by pressing the ‘Checkmark’
button. A login window will pop-up. An administrator login is required to store the new data set!