4 Description of function

18

NIRMaster Operation Manual, Version A

Factors that can influence the measurement:

Not enough sample material available

Sample not homogenious or representative

Humidity of sample material

Temperature of sample material

Sample cup inadeqate (e.g. cup material and thickness, blind spots, dirt etc.)

NOTE

The samples should either be measured at a constant temperature or the temperature influences

must be considered during calibration.
Coarse grained samples should be milled before carrying out the measurement. Recommended

number of scans: Qualitative calibration: 4−16; quantitative calibration: 16−32;
granules: 64.

4 .2 .2

Transflectance mode

Translucent and opaque liquids can be analyzed via transflectance mode. The light penetrates the
liquid, is diffusely reflected by the reference plate and passes the sample a second time. The trans-
flected rays contain the spectral information of the sample.

Factors that can influence the measurement:

Not enough sample material available

Air bubbles in the measurement path and under the transflectance cover

Sample not homogenious or representative

Temperature of sample material

Sample cup or transflectance cover inadequate (e.g. cup material and thickness, blind spots, dirt

etc.)

Diffuse reflecting / reference plate

Small, defined gap for liquid samples

Sensor

NIR light

Sample cup / petri dish

NOTE

The samples should either be measured at a constant temperature or the temperature influences

must be considered during calibration.
Recommended number of scans:

Qualitative calibration: 4−16; quantitative calibration: 16−32

Most liquid sample materials can be constantly analyzed with the help of a flow-cell adapter.