7 Maintenance and repairs

54

NIRMaster Operation Manual, Version A

7 .6 .2

Lamp exchange

The light source of the NIRMaster has a life expectancy of approx. 6000 hours. After this time, it
needs to be replaced. To avoid downtime, the NIRMaster is equipped with an independent secondary
backup lamp.

In case of a primary lamp module failure:

The system automatically switches to the secondary lamp in case of a primary lamp failure.

The secondary lamp will be in use until the primary lamp has been exchanged by the user or a

service person.
The operator will be reminded to exchange the primary lamp module. This should be done as soon

as possible.

!

CAUTION

Risk of minor burns by hot lamp module.

Do not touch hot parts or surfaces

Let the lamp module cool down safely

Do not touch the bulb glass

a

d

c

e

b