7 Maintenance and repairs
54
NIRMaster Operation Manual, Version A
7 .6 .2
Lamp exchange
The light source of the NIRMaster has a life expectancy of approx. 6000 hours. After this time, it
needs to be replaced. To avoid downtime, the NIRMaster is equipped with an independent secondary
backup lamp.
In case of a primary lamp module failure:
The system automatically switches to the secondary lamp in case of a primary lamp failure.
•
The secondary lamp will be in use until the primary lamp has been exchanged by the user or a
•
service person.
The operator will be reminded to exchange the primary lamp module. This should be done as soon
•
as possible.
!
CAUTION
Risk of minor burns by hot lamp module.
Do not touch hot parts or surfaces
•
Let the lamp module cool down safely
•
Do not touch the bulb glass
•