7 Maintenance and repairs

45

NIRMaster Operation Manual, Version A

7 .2 .2

External reference

For dependable reference measurements the External reference and Transflectance adapter must be
clean and damage free.

For cleaning, use abrasion-resistant precision cloths (e.g. Kimwipes from Kimberly-Clark). If
substances still adhere, you may additionally rinse the surfaces with alcohol.

NOTE
The External reference is sensitive to scratches. External referencing can only be performed manu-
ally via the Operator, not with the NIRMaster Operator software. Therefore for all NIRMaster Operator
applications the external referencing must be set to ‘interval’!

Do not drop the External reference!

Never bring it in contact with hard objects.

Do not use any cleaning media with abrasive character which might scratch the surface.

Do not use item for reference measurement when damaged or scratched.

7 .2 .3

Add-ons and custom sample vessels

Many sample vessels made of glass only, such as Petri dishes, can be cleaned in a washer with stan-
dard detergents. However, all BUCHI add-ons can be rinsed and cleaned with Meliseptol® liquid or
spray by B.Braun (disinfectants on alcohol basis).

Dishwasher-proof parts:

Petri dishes

Unbreakable cup

Spinner cpl.

XL add-on

High performance cup



NOTE
Spray-disinfection should be used in addition to surface disinfection with wipes only!

7 .3

How to perform a database backup

The operator is responsible for data backup. BUCHI recommends frequent backups with an advanced
backup strategy such as monthly, weekly and daily snapshots on independent data storages.