7 Maintenance and repairs
47
NIRMaster Operation Manual, Version A
7 .3 .2
Advanced selective archive actions
Via the entry ‘Archive Operations’ the NIRMaster Administrators can create a new, empty database
containing all subsettings i.e. security, user and lifecycle settings. A wizard will guide the Administrator
through the procedure. Applications and library calibrations can can be selected to be imported into
the new database. The old database will be transformed into a backup file.
7 .3 .2
Backup on USB storage devices
A simple variant for backups is to connect an external storage (e.g. external hard disk with USB 2.0 or
higher) to the NIRMaster.
NOTICE
Risk of instrument damage by liquid and particle ingress
Store and operate the instrument only in accordance to its ingress protection class only
•
Install dust caps and covers and housing parts tight to establish safe operation conditions
•
How to store a backup onto a USB storage device
Exit all NIRWare applications
•
Unscrew the dust cap
•
a from the NIRMaster USB socket
Connect the USB storage
•
b to the NIRMaster
Perform a backup as described in section 7.3.1 .
•
Copy the backup onto the USB storage.
•
Safely remove the USB storage from the NIRMaster and reinstall the dust cap
•
a
b
Backup
NOTE
Some USB devices are not bus powered and require an external power source. See storage user
•
manual for more information!
NIRWare administrator rights are required for backup.
•
When NIRAnywhere is installed and active, local backup functionality is disabled.
•