8 Troubleshooting

55

KjelFlex K-360 Operation Manual, Version C

Table 8-1: General malfunctions and their remedy (cont.)

Error number / Error indication

Possible cause

Corrective measure

35

There is no sample tube.

Sample tube present? Sample tube defec-
tive? Remove the sample tube and put it
back again. Restart the process

37

Boric acid level is low. Do you want to
continue determination?

Refill the boric acid tank

38

Water level for steam generator is low. Do
you want to continue determination?

Refill the water tank for the system
generator

39

Sodium hydroxide level is low. Do you want
to continue determination?

Refill the sodium hydroxide tank

40

Waste tank is full. Do you want to continue
determination?

Empty the waste tank

41

Titrator not ready. Retry?

Wait until the titrator is ready, try again

42

Not enough cooling water!

Cooling water connected? Cooling water
flowing? Flow sensor may be defective
Contact the Buchi customer service

43

Steam generator not ready.

Set status of steam generator to “Ready”
mode until the steam generator is ready
(up to 2 minutes)

44

Steam generator over temperature.

Check water connection

45

Cannot reach water level for steam
generator.

Fill water tank, check H

2

O pump

46

Titrator not ready or not configured.

Configure titrator in Configuration menu

47

Dispenser not ready or not configured.

Configure dispenser in Configuration menu

48

The printer is not ready for printing. The
current job will not be printed.

Switch on printer, check connection K-
360/printer

49

Printer: No paper.

Add paper and start printing again

50

The printer is busy. The current job will not
be printed.

Wait until the printer is ready and start
printing again

51

The system is offline. To set the system
online, go to Diagnostics.

Go to the Diagnostics menu and set the
instrument online

52

Unit runs in demo mode.

Switch off Demo mode in the menu
Configuration

53

Standby mode, switch to ready mode to
heat steam generator.

Switch to Ready mode

54

Determination still in progress.

Wait until the active determination is
finished and start again

55

Device cooled down! Priming or Preheating
recommended. Do you want to continue
determination?

Perform priming/preheating or continue
determination

56

Battery voltage low.

Call the Buchi service to replace the
battery

57

Unit supply voltage was too low! Please
switch off unit and start again.

Check the power supply source in your lab

58

An internal error occured! (Watchdog PCB)
Please switch off unit and start again.

Contact the Buchi customer service