6 after distillation 6.9, Switching on/off checklist, 6 .9 switching on/off checklist – BUCHI KjelFlex K-360 User Manual
Page 46
6 Operation
46
KjelFlex K-360 Operation Manual, Version C
6 .8 .6
After distillation
If you do not work with the titration set, empty the receiving vessel. The content of the receiving vessel
of the titration set is automatically aspirated at the end of the distillation, if this is configured within the
method.
It is recommended to clean the system using the cleaning method at the end of the day.
NOTE
The water volume can be adapted in the menu Cleaning under H2O.
6 .9
Switching ON/OFF checklist
To switch on and preheat the instrument, proceed as follows:
Check the filling level of the supply tanks.
•
Open the main water tap.
•
Switch on the instrument and wait until the steam generator is ready.
•
Insert an empty sample tube and an empty receiving vessel.
•
Close the protective door.
•
Start the preheating/priming by pressing the Start button in the corresponding menu.
•
Start the distillation by pressing the Start button.
•
To switch off the instrument, proceed as follows:
Start the cleaning process (see chapter 7.12).
•
Rinse the acid resistant pump with distilled water (in case it was used for the dosage of strong
•
acids, e.g. HCl).
Switch off the instrument.
•
Close the main water tap.
•