7 Maintenance and repairs

47

B-390

Operation Manual, Version C

7 .4

Cleaning

!

Warning

Pressure increase in the inlet-system due to clogged nozzles.

Bursting of the inlet system.

Death or serious poisoning by contact or incorporation of harmful substances at use.
• Clean nozzle immediately after use, see following section.

Wear laboratory coat

Wear protective goggles

Wear protective gloves

7 .4 .1

Cleaning the nozzle after each immobilization run

It is critical to clean the nozzle immediately after use so that the encapsulation medium (alginate, etc.)
will not dry and clog the system.

1. Leave the nozzle in place on the bead producing unit.

2. Attach a 20 mL or 60 mL syringe to the bead producing unit and inject 20 to 60 mL of distilled

water or of the solvent used for the encapsulation polymer.

3. If needed unscrew the nozzle from the bead producing unit, rinse with deionized water (see

figure 7-1) or an appropriate solvent and dry the nozzle with a flush of air.