beautypg.com

10 tips and tricks for a distillation, 1 what to do when the solvent starts foaming, 6 .10 tips and tricks for a distillation – BUCHI Vacuum Controller V-855 User Manual

Page 44

background image

6 Operation

44

V-850/855 Operation Manual, Version E

6 .10 Tips and tricks for a distillation

6 .10 .1

What to do when the solvent starts foaming?

• Press the button

P

. The current system pressure is frozen and the aeration valve is opened. The

system is aerated and switches into the Hold mode, i.e. the

H Off button becomes active.

• When the foam is gone or has sufficiently degenerated, press the

H Off button to continue with the

distillation process.

• In case the solvent starts foaming again, repeat the measures described above.

6 .10 .2

How to find out the distillation conditions for a solvent?

The solvent library integrated into the vacuum controller software provides you with information on a
number of different solvents.
To use the library, proceed as follows:
• Press the Library button. The solvent library screen opens.
• Select the solvent you are operating with by means of the selection knob on the vacuum controller

and press

OK.

• Now define the bath temperature you want to operate with by means of the selection knob on the

vacuum controller and press

Accept. The setting is saved and the value for pressure is adapted

accordingly.

NOTE
The values indicated for boiling point and cooling H

2

O max only serve for information purposes.

6 .10 .3

How to start a distillation without determining a pressure setpoint?

Preconditions:
• In the submenu

Remote control Evaporator, the options Rotation start and Immerse flask

have to be set to

Off.

For a distillation quick start, proceed as follows:
• On the main screen press the

Cont . button. The distillation starts.

• Now press the

Man . button to enter the Manual mode.

This manual is related to the following products: