beautypg.com

Brookfield Thermosel User Manual

Page 8

background image

Brookfield Engineering Laboratories, Inc.

Page 8

Manual No. M/94-204-I0612

Position the thermo-container and base between the viscometer stand legs and adjust the three (3)

knurled screws until the bubble is centered.

Lower the viscometer until the tips of the alignment bracket (see Figure I-1) just touch the horizontal

sur face of the locating ring, making contact directly behind the vertical curve. Raise the vis cometer,

positioning the tips of the alignment bracket about 1/16 inch above the horizontal surface behind

the locating ring. An etched line on the back of the locating ring is the 1/16 inch reference point.

SPINDLE

ALIGNMENT BRACKET

HT-69Y

LOCATING RING

Figure I-1

NOTE: Do not forcibly displace the alignment bracket!

Thermosels and Viscometers are factory-aligned prior to shipment. If viscosity measure ments are

suspect, first check system calibration using suitable viscosity standard fluids (see Appendix B).

Add the appropriate volume to the sample container (see Appendix A, Range Tables, for volume

requirement; see viscometer operating manual for calibration procedures).

• If using a spindle that requires an extension link and coupling nut, attach one end of the link to

the spindle and the other to the coupling nut. Carefully lower the spindle by the coupling nut

and link into the sample chamber. While pushing up gently on the pivot screw that protrudes

from the bottom of the viscometer, thread the coupling nut onto the viscometer. Note that the

spindle has a left-hand thread.

• If using a solid shaft (SC4-**BS) spindle, carefully lower the spindle into the sample chamber.

While pushing up gently on the spindle coupling nut that protrudes from the bottom of the

viscometer, thread the spindle onto the viscometer. Note that the spindles have a left-hand

thread.

Place the insulating cap over the sample chamber inlet.