I. INTRODUCTION

The Brookfield Thermosel System consists of a Brookfield Viscometer or Rheometer with related

accessories to accurately measure the viscosity of liquids at elevated tem peratures. Viscosity

measurements are obtained using coaxial cylinder geometry. Temperature is controlled by a solid

state, triac output, Proportioning Controller which maintains the spindle, chamber, and sample

material at the desired temperature.

The system is designed for measuring liquid viscosities over a temperature range from 40°C (104°F)

to 300°C (572°F). The designed viscosity range is from 3 cP to 80,000,000 cP depending upon the

vis cometer and SC4 spindle used.

Please note that all references made herein to digital viscometers would also apply to DV-III+

Rheometers and DV-III Ultra Rheometers.

I.1 Utilities

Input Voltage:

115 VAC or 230 VAC

Input Frequency:

50/60 Hz

Current Rating:

15 Amps

Main supply voltage fluctuations are not to exceed ± 10% of the nominal supply voltage.

I.2 Specifications

Operating Environment: 0°C to 40°C temperature range (32°F to 104°F)

Relative Humidity:

20% - 80%: non-condensing atmosphere

CE Certification:

This product conforms to the general safety requirements for

electrical equipment, for measurement control and laboratory use.

This product conforms to international standards which specify

minimum requirements for immunity and emissions regarding

electromagnetic compatibility (EMC) for electrical equipment