APPENDIX C — DISPOSABLE SAMPLE CHAMBERS AND SPINDLES

Disposable Sample Chambers are intended for use with difficult to clean materials, or in busy

laboratories where the low cost disposable chamber may improve productivity. Also available are

disposable SC4-27D spindles for use with the Brookfield Thermosel.

The HT-2DB disposable chambers are inserted into or removed from HT-60DP or HT-60ADP

Thermo Container using the HT-116 pliers as shown in Figure C-1.

Thermocontainer HT-60DP/HT-60ADP

Extractor Pliers

HT-116

Disposable Sample

Chamber HT-2DB

SC4 Series

Thermosel Spindle

Figure C-1

Replacement disposable chambers are available in packages of 100 chambers (Brookfield Part Number

HT-2DB-100). Contact Brookfield or your Brookfield agent for pricing and delivery information.

A package of five disposable aluminum sample chambers (Brookfield Part Number HT-2DB)

and one pair of extraction pliers (Brookfield Part Number HT-116) are supplied with each new

Thermosel System.

Disposable Spindle Instructions

(see Figure C-2)

1. Loosen closer.

2. Slide disposable spindle into chuck to insertion mark.

3. Hold chuck and hand-tighten closer.

4. Attach the assembly (spindle and chuck) to the viscometer being careful to hold the pointer

shaft firmly and lift up slightly.