beautypg.com

Brookfield Bath, Model TC-101 User Manual

Page 3

background image

Section 1. General Information

1.1

Unpacking

Your circulator is shipped in a special carton.

Retain the carton and all packing materials until the unit is

completely assembled and working properly. Set up and run the unit immediately to confirm proper operation.
Beyond one week, your unit may be warranty repaired, but not replaced. If the unit is damaged or does not
operate properly, contact the transportation company, file a damage claim and contact the company where your
unit was purchased.

1.2

Package Contents

All

TC-101P

TC-201P

TC-501P

Circulator Bath

Operators Manual

110-119

RS-232 Communication Cable

225-173

3

/

16

in.,

1

/

4

in., and

3

/

8

in. Nylon Barbed Tubing Adapters

510-011

6 ft. of

1

/

4

in. ID Latex Tubing

300-299

Beaker Platform(s) for Bath Reservoir

— 600 ml

701-402

701-402

(qty 2)

701-402

— 1000ml

701-403

Deck Lid(s)

— solid

510-209

510-209

— w/beaker holes

510-211

510-209

510-211

Blue Hole Plugs

3

1

/

2

inch

300-295

300-295

(qty 2)

300-295

Note: Do

NOT

Use Above 120°C

4

1

/

4

inch

300-296

Note:

Work area "opening" is designed to measure samples directly in the bath. If additional viscometer height

is required (spindle/guard clearance), either a 4 inch rod extension (part number BLM-4E) used with type A lab
stand or an 18 inch rod replacement

(part mumber VS-38) used with type S lab stand are available from

Brookfield or an authorized dealer.

1.3

Description of Circulating Baths

The PROGRAMMABLE circulating baths are designed to provide precise temperature control of fluids for
circulation to external equipment or to be used as a stand alone bath. The reservoir may be used for immersing
samples while the unit is connected to an external device. All wetted parts are corrosion resistant 300 series
stainless steel. Models are equipped with various size reservoirs and refrigeration capacities.

1.4

Specification Chart

Controller

Programmable

Temperature Range (Non-Refrigerated)

Ambient +5° to 200°C

Temperature Range (Refrigerated)

-20° to 200°C

Temperature Stability

±.01°C

Readout

LCD

Readout Accuracy

±0.25°C

Pump

Duplex

Pump Flow Rate (Pressure) @ 120V, 60Hz

11 to 24 lpm

Pump Flow Rate (Suction)

8 to 18 lpm

Overtemp/Safety Cutoff

Yes (Adjustable)

RS232 Interface

Yes

Remote Probe

Optional

Inlet / Outlet

1

/

4

inch NPT (female)

Heater

1000 Watts

4